Electrical Specifications
-Voltage DC: 6.000V - 1000V -Voltage AC: 600.0 mV-1000V -Amps AC: 60.00 mA – 10A -Amps DC: 60.00 mA – 10A -Resistance: 600.0 Ohm - 50 M Ohm -Capacitance: 1nF- 9,999 µF -Frequency: 99.99Hz- 99.99kHz
|
General Specifications
-Safety: ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Measurement Category III, 600 V Measurement Category IV -Certifications: CSA, TÜV (EN61010), UL, ?, ; (N10140),VDE -Maximum voltage between any terminal and earth ground: 1000V -Surge protection: 8 kV peak per IEC 61010 -Ohm Fuse for mA inputs: 440 mA, 1000 V FAST Fuse -Ohm Fuse for A inputs: 11 A, 1000 V FAST Fuse -Display: Digital: 6000 counts, updates 4/sec Bar Graph: 33 segments; Updates 32/sec Frequency: 10,000 counts Capacitance: 1,000 counts -Electromagnetic Compatibility (EN 61326-1:1997): In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: specified accuracy ± 5 °C (9 °F)
|
Brand SKU
FLUKE-77-4
|
Rashi Part SKU
FLUKE-77-4
|
Warranty
Limited lifetime
|